ANATEL PAT700 TOC Analyzer

Reliable TOC Monitoring for Pharmaceutical Water Systems

The ANATEL PAT700 Total Organic Carbon (TOC) Analyzer is engineered to support pharmaceutical manufacturing by ensuring accurate, compliant water quality testing. Ideal for Purified Water (PW) and Water for Injection (WFI) systems, and for Clean-In-Place processes, the PAT700 helps meet critical pharmacopeial standards including USP<643>, USP<645>, and EP 2.2.44

Key Features

  • 3-in-1 Measurement Measures TOC, Uncompensated Conductivity, and Temperature from a single analyzer.
  • Pharmacopeial Compliance Can be fully validated to meet USP <643> and EP2.2.44 requirements for TOC, and USP <645> for Conductivity.
  • Extended Service Intervals Requires calibration and servicing only once every 12 months.
  • Automated Sample Capture Automatically collects a water sample when a TOC excursion is detected for root-cause analysis.
  • Pre-Loaded SOPs with Visual Guidance Includes intuitive, step-by-step Standard Operating Procedures (SOPs) with on-screen pictures and prompts.

Benefits

  • Ensure Regulatory Compliance Maintain confidence in meeting global pharmacopeial requirements with validated performance.
  • Simplify Operations and Training Built-in SOPs reduce operator errors and accelerate onboarding of new users.
  • Minimize Downtime and Maintenance Costs Long calibration and service intervals lower the total cost of ownership.
  • Rapid Troubleshooting Immediate sample capture helps diagnose and resolve TOC anomalies quickly and effectively.
  • Compact, Integrated Solution Combines multiple measurements in one unit, saving space and simplifying system integration.


Explore the Full Potential of the ANATEL PAT700 TOC Analyzer:

Explore ANATEL PAT700 Models

Running the ANATEL PAT700

 

ANATEL PAT700 TOC Analyzer Product Specifications

Interface Protocols 3 x 4-20mA outputs │ 2 isolated digital inputs │ 4 x isolated digital outputs | Ethernet
Platform Anatel
Industry Standards 21 CFR Part 11, EP <2.2.44>, USP <643>, USP <645>
Calibration Intervals and Options 3-point TOC calibration; Conductivity calibration to USP<645>; System suitability to USP and EP
Flow Rate 60 to 300 mL/min Sample flow rate
Fluid Compatibility Purified water, Water for injection (WFI)
Concentration Range 0.5 ppb minimum to 2000 ppb maximum as Carbon
Online Operation Sample inlet pressure 10 to 100 psi (0.69 to 6.89 Bar)
Detection Differential conductivity
Reporting ppb TOC │ µS/cm
IP Rating IP46
Environmental Conditions Maximum altitude: 3000 m
Installation Category II
Temperature Range 1°C to 95°C sample range
Power Requirements 100/230VAC ± 10%, 65W, 50/60Hz
Resolution 0.1 ppb
Display Full-color LCD touchscreen
Operating Modes Online TOC; Online Conductivity; Auto-switching Main and Standby UV lamps for 12 months operation │ 4-bottle auto-sampler for grab sample analysis │ Excursion capture for root cause analysis support
Pollution Degree 2
Weight 13.6 kg(30 lb)
Width 59.7 cm(23.5 in)
Depth 22.9 cm(9 in)
Height 25.4 cm(10 in)
Color Blue
Package Quantity 1

Technical Documents